Lucid – COMING SOON

Applying Purific’s reliability engineering to Type 1 benchtop polishing unit applications.

Type 1 Ultrapure Water

PRODUCT FEATURES

Water Quality Specifications

Water TypeResistivity / ConductivityTOC (ppb)Bacteria (cfu/mL)Endotoxins (EU/mL)Flow Rate
Type I (Ultrapure)18.2 MΩ·cm / 0.055 µS/cm<3<0.01<0.0012 L/min
Type II (Pure)<1 µS/cm<30<0.013, 5, or 10 L/h
Type III ~0.25 MΩ·cm / ~4 µS/cm<200<1

Key Features

  • Purification Stages:
    • Pretreatment: Removes particles ≥5 µm, chlorine, colloids, and organic material.
    • Reverse Osmosis (RO): Eliminates 95–98% of dissolved inorganic salts, >99% of dissolved organic matter (PM >100 Da), and 99.95% of microorganisms and particles.
    • Deionization: Ion exchange resins reduce conductivity to ≤1 µS/cm.
    • Ultrapurification Module: Further reduces ionic contaminants.
    • Photo-oxidation Lamp: UV radiation at 254 nm (germicidal) and 185 nm (oxidizes organic compounds).
    • Ultrafiltration Module (GRUF version): Removes pyrogens, RNases, and DNases.
    • Final 0.22 µm Filter: Ensures bacterial count <0.01 cfu/mL.
  • Dispensing Options:
    • Manual (continuous)
    • Volume-controlled
    • Time-controlled
  • Monitoring & Control:
    • 4.3″ color touch screen interface.
    • Real-time monitoring of water quality parameters: conductivity, resistivity, temperature.
    • Alerts for consumable replacements and system anomalies.
  • Maintenance:
    • Quick-connect cartridges with anti-drip system.
    • Semi-automatic sanitization process.
    • Calibration using certified standards traceable to national standards.

Physical Specifications

  • Dimensions (H × W × D): 60 × 36 × 49 cm
  • Weight: 35 kg
  • Power Supply: 110–220 VAC, 50–60 Hz
  • Feed Water Requirements:
    • Pressure: 2–6 bar
    • Temperature: ≤30°C
    • Hardness: ≤300 ppm (CaCO₃)
    • Conductivity: ≤1500 µS/cm
    • Free Chlorine: <1 ppm
    • Turbidity: <1 NTU

This system is suitable for various laboratory applications, including analytical methods (HPLC, ICP-MS, TOC analysis), molecular biology, cell culture, and preparation of reagents and buffers.


Reach out to register your interest at info@purific.com